DEVELOPMENT AND VALIDATION OF STABILITY INDICATING RP- HPLC METHOD FOR ESTIMATION OF TEGOPRAZAN IN API
Authors: Patel J , PANDYA U, DHIMAR M, PATEL D AND PRAJAPATI M

ABSTRACT
A simple, accurate, and sensitive RP-HPLC method was developed and validated for the estimation of Tegoprazan in API. The chromatographic separation was carried out using an Agilent eclipse ODS (150 × 4.6 mm, 5 ?m) column under isocratic conditions. The mobile phase consisted of acetate buffer and acetonitrile in an optimized ratio. The flow rate was maintained at 1.0 mL/min, with detection performed at 245 nm. The injection volume was set at 20 ?L, and the total run time was 30 minutes. Forced degradation studies were performed to evaluate the stability- indicating capability of the method were subjected to various stress conditions, including acidic, alkaline, oxidative, thermal, photolytic, and hydrolytic degradation. The developed method was validated according to ICH Q2 (R2) guidelines by assessing parameters such as specificity,linearity, accuracy, precision, limit of detection (LOD), limit of quantification (LOQ), and robustness. Linearity was established in the concentration range 150-450 ?g/ml for Tegoprazan. Hence, the developed RP-HPLC Method was found to be Specific, Accurate, Precise, and Robust. The result confirmed that the method is reliable, reproducible, and suitable for routine analysis. Keywords: Stability Indicating method, Forced Degradation, Method Development, Tegoprazan
Publication date: 01/12/2026
    https://www.ijbpas.com/pdf/2026/December/MS_IJBPAS_2026_10613.pdf
Download PDF
https://doi.org/10.31032/IJBPAS/2026/15.12.10613